Engineering Design Portfolio
RF Hand Piece for the Treatment of Dry Eye
Device Purpose
Used on the eyelids for the treatment of dry eye symptoms
Device Features
Hand piece designed for use with existing, on the market, RF generator
Designed for ultra-fast temperature response – important for safety when treating on eyelid
Final design currently on the market
My key contributions
Full mechanical design (from prototype to manufacturing) of hand piece i
Industrial design
Plastic part design
Low mass stainless tip and temperature sensor topology
Low noise thermistor flex PCB design
Spring plunger design for RF contact safety
Translation of ophthalmologist feedback on other devices to create an optimal design for this treatment
Integrating existing control PCB to reduce time to market
Prototype internal design with repurposed PCBs from other customer offerings